Problem of detection
Content by : Bastien Bommelaer,
Approved by : Antoine
Last update : 12/01/2022
Problem of detection
Issue:
It can happen that sometimes a i700 is not recognized by Medit Link on Dell PC (7760 for instance). This can be because the i700 is using a port that is already in use.
On Device manager, there are two items at Port (COM & LPT) and they have same port number (COM3).
Selected same port may affect to connect i700 scanner.
See below how to fix it.
How to fix:
- Open the device manager
- Right click on Medit Controller, choose the "Properties".
- Select 'Port Settings' tab and 'Advanced…' button.
- On 'COM Port Number', choose a port number not in use (8 in the screenshot). Then click on 'OK'
- Right click on Medit Controller, choose the 'Disable Device'
- Right click on Medit Controller again, choose the 'Enable Device'
- Then i700 Scanner will be connected.