In case of issue with the calibration pnl file server which avoid to perform the calibration of the scanner in Medit T-scan as bellow:

  1. You can use the following link to manually download the calibration file (PNL file): https://pnl.meditlink.cn/PNLDown/
  2. Here enter the serial number of the scanner (1), click on "Search" (2) and on "PNL File Download" (3). 
  3. Save the file somewhere on the system. 

  4. Return in T-Scan software and click on the calibration icon

  5. Click on Next

  6. Choose the manual option and click on Next

  7. Click on the folder icon, select the PNL file you downloaded previously and click on Next

  8. Click on Next to start the calibration